Technical Data 53
11.1 Safe distances
The ACCUVAC Basic is intended for operation in an
electromagnetic environment in which high-fre-
quency interference variables are controlled. The
customer or user of the ACCUVAC Basic can help
avoid electromagnetic interference by maintaining
the minimum safe distance between portable and
mobile high-frequency telecommunication devices
(transmitters) and the ACCUVAC Basic depending
on the rated output of the transmitter as given be-
low.
Recommended safe distances between portable or mobile HF
telecommunication devices (e.g. mobile phones) and the ACCUVAC Basic
Rated output of
HF device
in W
Safe distance depending on transmission frequency
in m
150 kHz - 80 MHz
d=(3,5/V1) x √P
80 MHz - 800 MHz
d=(3,5/V1) x √P
800 MHz – 2,5 GHz
d=(3,5/V1) x √P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33