Functional check 39
EN
forcement clip on the latch (reusable collec-
tion canisters)
tube connections and disposable collection
bag (disposable collection canisters).
3. Switch on ACCUVAC Basic.
4. Check the charge level on the voltage indicator.
Always recharge the power pack without delay
if the yellow LED on the voltage indicator lights
up (see “5.4 Charging ACCUVAC Basic” on
page 26).
5. Insert the stopper in the fingertip.
6. Use your thumb to hold the suction nozzle
closed.
7. Switch on the aspirator and select the maximum
vacuum of –0.8 bar by turning vacuum control
fully to the right. ACCUVAC Basic must reach
this vacuum in not more than 20 seconds.
If the aspirator takes more than 20 seconds to
reach this vacuum, its suction capacity is re-
duced. Check for possible faults (see “8. Trou-
bleshooting” on page 40).
8. Switch ACCUVAC Basic off again.