44 Programming Welch Allyn AED 10 Automatic External Defibrillator
Accessing the Supervisor Menu
The Supervisor Menu is accessed from the Main Menu.
To access the Supervisor Menu
1. While in the Main Menu, press the Next soft key to highlight Supervisor, and then
press the Enter soft key to go to the Password Menu.
2. Enter the Password by using Next to scroll to the appropriate number and use Enter
to shift to the next digit. After the third number has been selected, highlight Save by
pressing Enter.
3. Press Enter again while in the Save box. If the password is correct, the Supervisor
Menu will appear. If the password is incorrect, the Main Menu will appear.
4. To exit the Password prompt, press Next when the Save box is highlighted to select
the Cancel box. Press the Enter soft key to return to the Main Menu.
Supervisor Menu Items
Note
The default supervisor password is 1 2 3.
Volume
Contrast
Date
Time
Log
Supervisor
Main Menu
Password
Save
Cancel
000
Password
Save
Cancel
123
Protocol
Shock Mode
Program
More
Back
CPR
Supervisor
Language
CPR Set the value for CPR Timer and/or CPR Tempo.
Language Select text and audio language. The Welch Allyn AED 10 will
reboot into automated mode to activate selected language.
Protocol Select the energy level protocol. Default protocol is 200, 300,
360 Joules.
Shock Mode Selects among 1 Shock CPR (default), 3 Shock CPR, or 3 Shock
Analyze mode.
Protocol
Shock Mode
Program
More
Back
CPR
Supervisor (1)
Language
Dept
Change Code
Bit Rate
Utility
Back
Restore Defaults
Supervisor Menu
Unit ID