68 Specifications Welch Allyn AED 10 Automatic External Defibrillator
Pediatric Energy Reducer
Display
Battery
Energy Attenuation Factor 25%, at 25 Ω, 50 Ω and 75 Ω
20%, at 100 Ω and 125 Ω
15%, at 150 Ω and 175 Ω
Type Backlit Monochrome LCD
Size 2.1” x 1.42” (54 x 36 mm)
Resolution 160 x 100
Freq Response 1 to 40 Hz
Low Battery Indicator Battery Icon gauge on display with 10 capacity levels
Backlight LED
Type Non-rechargeable, LiMnO2 12V, 2.6 Ah
Capacity 90 discharges at 360 Joules or
125 discharges at 200 Joules or
150 discharges at 150 Joules or
5 hours ECG monitoring
Shelf Life
25°C ± 15°C
10 years (5 years storage + 5 years standby)
5 years standby (after installation)
Battery capacity measured according to IEC 60601-1-2-4, claf
use 102.3.2 at room temperature. Capacity may be diminished by extremes, or when