EVOM™ Manual
World Precision Instruments 33
Fig. 63—(Left) When the battery level falls to 1%, you are asked to save your data le.
Fig. 64—(Right) Once the data is save, this message appears.
When the battery level falls to 0%, the system begins a 60 second countdown before
powering o.
Fig. 65—When the battery power falls to 0%, a warning displays briey.
Fig. 66—The 60 second countdown was initiated when the battery power fell to 0%.
Before shutdown, a nal warning appears.
Fig. 67—Immediately before the system powers o, the shutdown message displays.
Reloading a Backup File after Power Failure
NOTE: This feature is only available in the USB save mode. When using the EVOM™
Companion Application, individual .CSV les are created for each plate in memory
and are in the same folder where all other .CSV les are saved using the le naming
convention designated on the Store Options screen.
1. When power is restored a green message appears (Fig. 68). Press
OK
.
Fig. 68—(Left) Power level is restored when the unit is plugged into a power source.
2. Cycle power on the EVOM™ Manual by turning it o and on again by sliding the
power switch on the back panel. This automatically loads the backup les which

Related product manuals