2 World Precision Instruments
Repeat the sampling process. Press the
Preview
button to see the measurements
that have been taken.
6. Once the last well has been recorded, the le storing selection notication appears
prompting you to take an action.
Select one of the options as follows:
Export
starts the storage of all the plate les. When you select it a message
appears describing which plate is being stored. The number of plates
could be one, two or three, depending on how many were collected before
exporting.
Next
returns you to the main screen so you can take measurements of the
next plate which starts from a clear state.
NOTE: If you are on plate 3, the
Next
button does not appear, because EVOM™
Manual’s memory only holds data from three plates.
Back
returns you to the main screen (data collection). You go back to the rst
well of the current plate being collected and can acquire the data again.
7. When the plate data has been saved, a conrmation message appears. Press the
Done
button to return to the main screen.
8. Now you can work with your data.
If you stored your data on the USB ash drive, remove the drive and transfer
the data les to your computer. They are in comma separated value (CSV) le
format for Microsoft
®
Excel.
If you stored your data using the EVOM™ Companion Application, all your
data les will be located in the directory you specied when you started the
Companion Application.
TEER Calculations
To compute TEER, multiply the measured resistance by the surface area’s listed below.
For example, a 12 mm insert measures 565 Ω, the TEER is 565Ω * 1.13 or 638 Ω-cm
2
.
6 well plate (24 mm inserts) 4.52 cm
2
12 well plate (12 mm inserts) 1.13 cm
2
24 well plate 6.5 mm inserts) 0.3316 cm
2
96 well plate (4.3 mm inserts) 0.145 cm
2