EVOM™ Manual
World Precision Instruments 39
Issue
Possible Cause Solution
The reading appears to be
drifting* (see the drift denition
below).
The electrodes
are coated with
deposits from the
use of media or
buer.
The electrodes require enzymatic cleaning.
In a 5% CO
2
environment, a
loss of CO
2
causes
the media pH to
change, and the
resistance reading
may increase
sharply.
A 5% CO
2
environment can help in reducing pH media
changes.
The electrode
reading is unstable.
(See the stability
denition below).
The electrodes are
coated.
The electrodes require enzymatic cleaning.
Radio frequency
interference
Turn o or move any cellular phones farther away from the
experimental setup.
---- appears on
the screen
Resistance reading
is out of range
An out of resistance range will display ---- on the screen. Adjust
the range to correct the error. A saved well will read a space
(“ ") if the display is out of range. Go to
Setup
,
Mode Units
and
increase the resistance range. See page 11.
NOTE: The EVOM™ Manual will read only about 5% above
100K Ω.
NOTE: If you have a problem/issue with that falls outside the denitions of this
troubleshooting section, contact the WPI Technical Support team at (941) 371-1003 or
wpi-ts@wpiinc.com.
*Drift–Readings that continuously increase or decreases a signicant value (either
voltage or resistance) over time. Example: At 1000 Ω the reading is increasing 100
Ω/minute. (A drift of 10 Ω/minute is acceptable.) Excessive drift may be caused by
unwanted deposits on the electrode, pH changes or temperature variance.
Instability–At 500 Ω, the reading jumps from 450 to 550 Ω and will not settle down (an
instability ±5 Ω is acceptable in the 500 Ω range. In the higher ranges, up to ±1000Ω is
acceptable at the 100K range. Electrodes showing instability may be coated with media
or buer constituents and may require enzymatic cleaning.