EasyManua.ls Logo

WPI EVOM - Page 43

WPI EVOM
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EVOM™ Manual
World Precision Instruments 39
Issue
Possible Cause Solution
The reading appears to be
drifting* (see the drift denition
below).
The electrodes
are coated with
deposits from the
use of media or
buer.
The electrodes require enzymatic cleaning.
In a 5% CO
2
environment, a
loss of CO
2
causes
the media pH to
change, and the
resistance reading
may increase
sharply.
A 5% CO
2
environment can help in reducing pH media
changes.
The electrode
reading is unstable.
(See the stability
denition below).
The electrodes are
coated.
The electrodes require enzymatic cleaning.
Radio frequency
interference
Turn o or move any cellular phones farther away from the
experimental setup.
---- appears on
the screen
Resistance reading
is out of range
An out of resistance range will display ---- on the screen. Adjust
the range to correct the error. A saved well will read a space
(“ ") if the display is out of range. Go to
Setup
,
Mode Units
and
increase the resistance range. See page 11.
NOTE: The EVOM™ Manual will read only about 5% above
100K Ω.
NOTE: If you have a problem/issue with that falls outside the denitions of this
troubleshooting section, contact the WPI Technical Support team at (941) 371-1003 or
wpi-ts@wpiinc.com.
*Drift–Readings that continuously increase or decreases a signicant value (either
voltage or resistance) over time. Example: At 1000 Ω the reading is increasing 100
Ω/minute. (A drift of 10 Ω/minute is acceptable.) Excessive drift may be caused by
unwanted deposits on the electrode, pH changes or temperature variance.
Instability–At 500 Ω, the reading jumps from 450 to 550 Ω and will not settle down (an
instability ±5 Ω is acceptable in the 500 Ω range. In the higher ranges, up to ±1000Ω is
acceptable at the 100K range. Electrodes showing instability may be coated with media
or buer constituents and may require enzymatic cleaning.

Related product manuals