EasyManua.ls Logo

wtw Cond 3210 - Displaying Calibration Records

wtw Cond 3210
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Cond 3210
36
ba75794e04 04/2012
Determining the cell
constant
For this calibration procedure, the Measuring cell setting must be set to
cal in the measuring menu. Proceed as follows to determine the cell
constant:
4.4.6 Displaying calibration records
Displaying the
calibration record
The calibration record of the last calibration is to be found under the
menu item, Calibration / Calibration record. To open it in the measured
value display, press the <CAL
_> key.
The calibration records of the last calibrations (up to 10) are available
in the menu, <F1>/[Menu] / Calibration / Calibration data storage and
in the menu, <F1
_>/[Menu] / Storage & config/Data storage / Calibra-
tion data storage.
1 Connect a conductivity measuring cell to the measuring
instrument.
2 In the measured value display, select the conductivity
parameter with <M>.
3 Start the calibration with <CAL>.
The cell constant that was calibrated last is displayed.
4 Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
5 Start the measurement with <ENTER>.
The [AR] status indicator is displayed. The measured
parameter flashes.
6 Wait for the end of the AutoRead measurement or accept the
calibration value with <ENTER>.
The calibration record is displayed.
7Using <F1>/[Continue] or <ENTER>, switch to the measured
value display.
ӕ
0.839
1/cm
25.1
°C
[Tr25][nLF][ 0.835 1/cm]
22.09.2008
08:00
Menu

Related product manuals