Operation Cond 3210
42
ba75794e04 04/2012
4.6 Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.6.1 Resetting the measurement settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
The following settings for conductivity measurements are reset to the
default settings with the Reset function:
The sensor settings are reset under the Reset menu item in the
measuring menu. To open the settings, activate the relevant measuring
window in the measured value display and press the <F1>[Menu] key
shortly
.
Setting Default settings
Cal. interval 150 d
Measured parameter ӕ
Cell constant (c) 0.475 cm
-1
(calibrated)
0.475 cm
-1
(set up)
Temperature compensation nLF
Reference temperature 25 °C
Temperature coefficient (TC) of
the linear temperature
compensation
2.000 %/K
TDS factor 1.00
Stability control On
Temperature unit °C