Operation Multi 197i
22
ba75339e04 07/2009
Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.8.2
before calibrating. After a valid calibration, the record is printed.
Sample printout:
CALIBRATION PROTOCOL
02.03.02 14:19
Device No.: 12345678
Calibration pH
Cal time: 01.03.01 / 15:20
Cal interval: 7d
AutoCal TEC Tauto
Buffer 1 2.00
Buffer 2 4.01
Buffer 3 7.00 *
Buffer 4 10.01
C1 184.1 mV 25.0°C
C2 3.0 mV 25.0°C
S1 -59.4 mV/pH
ASY1 - 4 mV
Probe: +++