EasyManuals Logo

wtw Multi 197i User Manual

wtw Multi 197i
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
Multi 197i Operation
35
ba75339e04 07/2009
4.4.6 Entering the salt content (salinity)
A salt content correction is required in the oxygen concentration mea-
surement of samples with a salt content of more than 1 g/l. To do this,
you have to enter the salinity equivalent (the measured salinity) of the
test sample (range 0.0 - 70.0) and to switch on the salinity correction.
Note
With the Multi 197i, you can measure the salinity. How to proceed is de-
scribed in section 4.5.3 M
EASURING THE SALINITY on page 38.
Entering the
salt content
Note
How to switch on the salt content correction is described on page 29.
1 Determine the salinity of the test sample (any method, see also
section 4.5.3 M
EASURING THE SALINITY on page 38).
2 Press the <CAL> key repeatedly until Sal appears on the dis-
play.
3 Enter the salt content with <> <>.
4 Switch to the measuring mode with <M>.
TP
Oxi
Cal
AR
°
C
S
Year
Ident
No.
Day.Month
Tref25 Tref20
LoBat
Lin
nLF
Sal
pH
1
TEC
Auto
Store
8
Time Baud
cm
1/
K
/
%
RCL
REL2
pH/mV
m
cm
S/
%
5
6 9 6
50
REL1
O
mg/l
Sal
ARng

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 197i and is the answer not in the manual?

wtw Multi 197i Specifications

General IconGeneral
Brandwtw
ModelMulti 197i
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals