Multi 197i What to do if...
67
ba75339e04 07/2009
6 What to do if...
6.1 pH system messages
Error message 0FL
Error message E3
Cause Remedy
pH electrode:
Not connected Connect electrode
Air bubble in front of the
diaphragm
Remove air bubble
Air in the diaphragm Extract air or
moisten diaphragm
Cable broken Replace electrode
Gel electrolyte dried out Replace electrode
Cause Remedy
Electrode
Diaphragm contaminated Clean diaphragm
Membrane contaminated Clean membrane
Moisture in the plug Dry plug
Electrolyte obsolete Replenish electrolyte or
replace electrode
Electrode obsolete Replace electrode
Electrode broken Replace electrode
Measuring instrument:
Incorrect calibration procedure Select correct procedure
Incorrect solution temperature
(without temperature sensor)
Set up correct temperature
Socket damp Dry socket