EasyManua.ls Logo

wtw Multi 197i - Dissolved Oxygen

wtw Multi 197i
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multi 197i Technical data
75
ba75339e04 07/2009
7.2.2 Dissolved oxygen
Measuring ranges
and resolutions
Variable Measuring range Resolution
Concentration [mg/l] 0 ... 19.99
0 ... 90.0
0.01
0.1
Saturation [%] 0 ... 199.9
0 ... 600
0.1
1
T [°C] 0 ... 50.0 0.1
Accuracies Variable Accuracy (± 1 Digit)
Concentration [mg/l] ± 0.5 % of measured value
at ambient temperature of 5 °C ... 30 °C
Saturation [%] ± 0.5 % of measured value
when measuring in the range of ± 10 K
around the calibration temperature
T [°C] ± 0.1
Correction
functions
Temperature com-
pensation
Accuracy better than 2 % at 0 ... 40 °C
Salinity correction 0 ... 70.0 SAL
Air pressure
correction
Automatic through integrated pressure sen-
sor in the range of 500 ... 1100 mbar