EasyManua.ls Logo

wtw Multi 197i - Page 25

wtw Multi 197i
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multi 197i Operation
25
ba75339e04 07/2009
Note
At this point, the AutoCal TEC calibration can be terminated with <M>.
This corresponds to a single-point calibration. To do this, the instru-
ment uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the
asymmetry of the electrode.
5 When the measured value is stable, Ct2 appears.
6 Thoroughly rinse the electrode with distilled water.
7 If required, enter the temperature of the second buffer solution
with <> <> while keeping the <RUN/ENTER> key de-
pressed.
8 Immerse the pH electrode into the second buffer solution.
9 Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example:
O
mV
S
/
1
Sal
O
LoBat nLF
Lin
Oxi
Tref25 Tref20
ARng
8
Time
Year
Ident
No.
Day.Month
Baud
Sal
cm
1/
K
/
%
RCL
REL2
pH
m
cm
S/
%
mg/l
2 4 8
°
C
TP
REL1
AR
Cal
TEC
Auto
/
1
Sal
S
69
LoBat nLF
Lin
Oxi
Tref25 Tref20
8
Time
Year
Ident
No.
Day.Month
Baud
Sal
cm
1/
K
/
%
RCL
REL2
pH
m
cm
S/
%
mg/l
2 4 8
3
°
C
TP
REL1
mV
AR
Cal
TEC
Auto
ARng
pH
700
pH
Store
AR
1
LoBat nLF
Lin
Oxi
Tref25 Tref20
ARng
8
Time
Year
Ident
No.
Day.Month
Baud
Sal
cm
1/
K
/
%
RCL
REL2
pH/mV
m
cm
S/
%
mg/l
Sal
O
S
pH
C
2 4 8
T2
°
C
TP
REL1
Cal
TEC
Auto

Related product manuals