Multi 197i Operation
51
ba75339e04 07/2009
2 Press the <RUN/ENTER> key.
The data of the last calibration of all measured variables ap-
pears in the following sequence:
z pH: Slope SLO and asymmetry ASY
z Oxygen: Relative slope SLO
z Cond: Cell constant C
Information concerning the calibration procedure is output as
well.
3 Press <RUN/ENTER> to display the value of the asymmetry
(mV).
4 Press <RUN/ENTER> to display the relative slope of the D. O.
sensor.