Multi 197i Operation
53
ba75339e04 07/2009
Outputting the calibra-
tion protocol on the
interface
Sample printout:
1 Press the <RCL> key repeatedly until CAL SEr appears on the
display.
2 Press the <RUN/ENTER> key.
The calibration protocol for all measured variables is transmit-
ted to the interface. After the data transmission, the instrument
automatically switches to the last active measurement mode.
d
Sal
mg/l
ARng
O
TP
Oxi
Cal
AR
°
C
S
Year
Ident
No.
Day.Month
Tref25 Tref20
LoBat
Lin
nLF
Sal
pH
C
TEC
Auto
Store
Time Baud
cm
1/
K
/
%
REL2
pH/mV
m
cm
S/
%
C
S E R
AL
REL1
RCL
CALIBRATION PROTOCOL
02.03.02 14:19
Device No.: 12345678
Calibration pH
Cal time: 01.03.02 / 15:20
Cal interval: 7d
AutoCal TEC Tauto
Buffer 1 2.00
Buffer 2 4.01
Buffer 3 7.00 *
Buffer 4 10.01
C1 174.1 mV 25.0°C
C2 -133.3 mV 25.0°C
S1 -59.4 mV/pH
ASY1 - 4 mV
Probe: +++
CALIBRATION 02
Cal time: 02.03.02 / 14:19
Cal interval: 14d
OxiCal Tauto AR
Relative Slope: 0,88
Probe: +++
Calibration Conductivity
Cal time: 14.01.02 / 11:37
Cal interval: 180d
Cal Std.: 0.01 mol/l KCL
40.0 °C
Conduct./Tref25: 1413µS/cm
Cell Const : 0.478 1/cm
Probe : +++