EasyManua.ls Logo

wtw pH 7310 - Page 26

wtw pH 7310
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
pH value pH 7310
26 ba75929e02 10/2011
Continuing with two-
point calibration
5. Thoroughly rinse the electrode with deionized water.
6. Immerse the electrode in the first buffer solution.
7. When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
8. Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
9. Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
10. If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the elec-
trode.
pH 1
-180.0
24.8 °C
AutoCal TEC
01.07.2011
08:00
Buffer
mV
pH
-180.0
24.8 °C
AutoCal TEC
10.011
01.07.2011
08:00
AR
Buffer
mV
11. Thoroughly rinse the electrode with deionized water.
12. Immerse the electrode in the second buffer solution.
13. When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.

Table of Contents

Related product manuals