pH 7310 What to do if...
ba75929e02 10/2011 61
12.2 General information
Sensor symbol
flashes
Display
Instrument does not
react to keystroke
You want to know
which software
version is in the
meter
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
Batteries almost empty Replace the batteries (see
section 11.1 MAINTENANCE,
page 57)
Cause Remedy
Operating condition undefined or
EMC load unallowed
Processor reset:
Press the <ENTER> and <On/
Off> key simultaneously
Cause Remedy
E. g., a question by the service
department
Switch on the meter.
Open the menu, <F1_>/
[Menu] / Storage & config /
SystemService information.
The instrument data are dis
-
played.