pH 7310 Settings
ba75929e02 10/2011 43
7.3 Sensor-independent settings
7.3.1 System
To open the Storage & config menu, press the <F1_>[Menu] key in the mea-
sured value display. After completing the settings, switch to the measured
value display with <M>.
Default settings are printed in bold.
Resolution mV 0.1
1
Resolution of the mV display
Stability control On
Off
Switches on or off the automatic stability
control during measurement (see
sec-
tion Stability control (AutoRead) &
HOLD function, page 22 )
Reset - Resets all sensor settings to the delivery
condition (see
section 7.4.1 RESETTING
THE MEASUREMENT SETTINGS, page 44).
Menu item Possible set-
ting
Description
System / General /
Language
Deutsch
English
(further)
Selects the menu language
System / General /
Beep
On
Off
Switches on/off the beep on
keystroke
System / General /
Illumination
Auto
On
Off
Switches the display illumina-
tion on/off
System / General /
Contrast
0 ... 50 ... 100 % Changing the display contrast
System / General /
Switchoff time
10 min ... 1h ...
24 h
Adjusts the switch-off time
System / Interface /
Baud rate
1200, 2400,
4800, 9600,
19200
Baud rate of the data interface
System / Interface /
Output format
ASCII
CSV
Output format for data trans-
mission (see section 9 TRANS-
MITTING DATA (USB INTERFACE),
page 52)
System / Interface /
Decimal separator
Dot (xx.x)
Comma (xx,x)
Decimal separator
Menu item Possible
setting
Description