53IM 04P02B01-02E
Setup Items and Default Values
Setup Items in Setting Mode and Their Default Values (Firmware version: 1.31)
The items with an asterisk are not displayed in the default condition. To display these
items, settings must be changed in Basic Setting mode.
Setup Item Pen/Dot Selectable Range or Selections Default Value
Range > CH - 01 to 24 01
Range > Mode - Volt/TC/RTD/1-5V/Scale Volt
/Delta/DI/SQRT/Skip
Range > Mode > Range - 20mV/60mV/200mV/2V/6V/20V/50V/1-5V 2V
/R/S/B/K/E/J/T/N/W/L/U/WRe/Pt/JPt/Level/Cont
/(Slections for /N1 and /N3 options)
*Bias > CH - 01 to 24 01
*Bias > Bias - On/Off Off
*Bias > Bias - ±10 % of the span of the measurable range -
Calibration correction (/CC1 option)
Setup Item Pen/Dot Selectable Range or Selections Default Value
Calibration > CH - 01 to 24 01
Calibration > Calibration - On/Off Off
Calibration > datums - Within the range -
Calibration > revise - Within the range -
(correction point + correction value for revise value)
Calibration > Decision - Yes/No Yes
Setup Item Pen/Dot Selectable Range or Selections Default Value
Alarm > CH - 01 to 24 01
Alarm > Level - 1/2/3/4 1
Alarm > Alarm - On/Off Off
Alarm > Type - H/L/h/l/R/r (*/T/t) H
Alarm > Value - Depends on the alarm type. -
Alarm > Relay - On/Off Off
Alarm > Relay number - I01 to I06, I11 to I16, I21 to I26, I31 to I36 I01
Unit > CH - 01 to 24 01
Unit > Unit - 6 characters or less Blank
Chart Pen Model 82 types (pen model) 25 mm/h
Dot Model 1 to 1500 mm/h (dot model) 25 mm/h
Clock - Date/Time -
Aux > Trend Dot Model Auto/Fix Auto
Aux > Zone > CH - 01 to 24 01
Aux > Zone > Left, Right - Within the recording span range (mm) Left: 0, Right: 180
*Aux > Partial > CH - 01 to 24 01
*Aux > Partial > Partial - On/Off Off
*Aux > Partial > Expand - 1 to 99% 50
*Aux > Partial > Boundary - Within the recording span range -
Aux > Print out > CH - 01 to 24 01
Aux > Print out > Trend Dot Model On/Off On
Aux > Print out > Periodic - On/Off On
Aux > Tag > CH - 01 to 24 01
Aux > Tag > Tag - 7 characters or less 01 to 24
Aux > Message > Message number - 1 to 5 1
Aux > Message > (Message) - 16 characters or less Blank
Aux > Chart2 Pen Model 82 types (pen model) 25 mm/h
Dot Model 1 to 1500 mm/h (dot model) 25 mm/h
*Aux > Moving_AVE > CH Dot Model 01 to 24 01
*Aux > Moving_AVE > number of samples Dot Model Off, 2 to 16 Off
*Aux > Filter > CH Pen Model 01 to 24 01
*Aux > Filter > Response time Pen Model 2s/5s/10s Off
*Aux > Alarm delay time > CH - 01 to 24 01
*Aux > Alarm delay time > Duration - 0 to 3600s 10s
Aux > Brightness > Display - 1/2/3/4/5/6/7/8 4
Aux > Brightness > Light - 1/2/3/4 2