7-63
IM AQ6370C-17EN
Remote Commands
1
2
3
4
5
6
7
8
App
Index
:CALCulate:PARameter[:CATegory]:SWPK
rms:TH
Function Sets/queries the threshold level of the PEAK-
RMS method–based spectrum width analysis
function.
Syntax
:CALCulate:PARameter[:CATegory]:SWP
Krms:TH<wsp><NRf>[DB]
:CALCulate:PARameter[:CATegory]:SWP
Krms:TH?
<NRf> = Threshold level [dB]
Example
:CALCULATE:PARAMETER:SWPKRMS:
TH 3.00db
:CALCULATE:PARAMETER:SWPKRMS:TH?->
+3.00000000E+000
Explanation This is a sequential command.
:CALCulate:PARameter[:CATegory]:SWRM
s:K
Function Sets/queries the magnification of the RMS
method–based spectrum width analysis
function.
Syntax
:CALCulate:PARameter[:CATegory]:SWR
MS:K<wsp><NRf>[DB]
:CALCulate:PARameter[:CATegory]:SWR
MS:K?
<NRf> = Magnification
Explanation
:CALCULATE:PARAMETER:SWRMS:K2.00
:CALCULATE:PARAMETER:SWRMS;K? ->
+2.00000000E+000
Explanation This is a sequential command.
:CALCulate:PARameter[:CATegory]:SWRM
s:TH
Function Sets/queries the threshold level of the RMS
method–based spectrum width analysis
function.
Syntax
:CALCulate:PARameter[:CATegory]:SWR
MS:TH<wsp><NRf>[DB]
:CALCulate:PARameter[:CATegory]:SWR
MS:TH?
<NRf> = Threshold level [dB]
Example
:CALCULATE:PARAMETER:SWRMS:
TH 3.00db
:CALCULATE:PARAMETER:SWRMS:TH?->
+3.00000000E+000
Explanation This is a sequential command.
:CALCulate:PARameter[:CATegory]:SWTH
Resh:K
Function Sets/queries the magnification of the THRESH
method–based spectrum width analysis
function.
Syntax
:CALCulate:PARameter[:CATegory]:SWT
HResh:K<wsp><NRf>
:CALCulate:PARameter[:CATegory]:SWT
HResh:K?
<NRf> = Magnification
Example
:CALCULATE:PARAMETER:SWTHRESH:
K 2.00
:CALCULATE:PARAMETER:SWTHRESH:K?->
+2.00000000E+000
Explanation This is a sequential command.
:CALCulate:PARameter[:CATegory]:SWTH
resh:MFIT
Function Sets/queries whether to enable the mode fit of
the THRESH method-based spectrum width
analysis function.
Syntax
:CALCulate:PARameter[:CATegory]:SWT
Hresh:MFIT<wsp>OFF|ON|0|1
:CALCulate:PARameter[:CATegory]:SWT
Hresh:MFIT?
Response 0 = OFF, 1 = ON
Example
:CALCULATE:PARAMETER:SWTHRESH:
MFIT ON
:CALCULATE:PARAMETER:SWTHRESH:MF
IT?-> 1
Explanation This is a sequential command.
:CALCulate:PARameter[:CATegory]:SWTH
resh:TH
Function Sets/queries the threshold level of the THRESH
method–based spectrum width analysis
function.
Syntax
:CALCulate:PARameter[:CATegory]:SWT
Hresh:TH<wsp><NRf>[DB]
:CALCulate:PARameter[:CATegory]:SWT
Hresh:TH?
<NRf> = Threshold level [dB]
Response ex. Same as above
Explanation
:CALCULATE:PARAMETER:SWTHRESH:
TH 3.00DB
:CALCULATE:PARAMETER:SWTHRESH:TH?->
+3.00000000E+000
Explanation This is a sequential command.
7.6 Instrument-Specific Commands

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA AQ6370C and is the answer not in the manual?

YOKOGAWA AQ6370C Specifications

General IconGeneral
BrandYOKOGAWA
ModelAQ6370C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals