EasyManua.ls Logo

YOKOGAWA DL1640

YOKOGAWA DL1640
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-23
IM 701610-61E
SPI-Bus Analysis Function
2
2.6 Communication Commands
:SEARch:SPI:ANALyze:SETup:CLOCk?
Function Queries all settings related to the clock channel of
the SPI analysis conditions.
Syntax :SEARch:SPI:ANALyze:SETup:CLOCk?
Example :SEARCH:SPI:ANALYZE:SETUP:CLOCK? ->
:SEARCH:SPI:ANALYZE:SETUP:CLOCK:
LEVEL 1.000E+00;HYSTERESIS 0.3;
POLARITY RISE
:SEARch:SPI:ANALyze:SETup:CLOCk:
HYSTeresis
Function Sets the clock channel hysteresis of the SPI
analysis conditions or queries the current setting.
Syntax :SEARch:SPI:ANALyze:SETup:CLOCk:
HYSTeresis {<NRf>}
:SEARch:SPI:ANALyze:SETup:CLOCk:
HYSTeresis?
<NRf> = 0.3 to 4.0 (div, 0.1 steps)
Example :SEARCH:SPI:ANALYZE:SETUP:CLOCK:
HYSTERESIS 0.5
:SEARCH:SPI:ANALYZE:SETUP:CLOCK:
HYSTERESIS? -> :SEARCH:SPI:ANALYZE:
SETUP:CLOCK:HYSTERESIS 0.5
:SEARch:SPI:ANALyze:SETup:CLOCk:
LEVel
Function Sets the clock channel level of the SPI analysis
conditions or queries the current setting.
Syntax :SEARch:SPI:ANALyze:SETup:CLOCk:
LEVel {<voltage>}
:SEARch:SPI:ANALyze:SETup:CLOCk:
LEVel?
<voltage>=8 divisions within the
screen (0.01 division steps).
Example :SEARCH:SPI:ANALYZE:SETUP:CLOCK:
LEVEL 1V
:SEARCH:SPI:ANALYZE:SETUP:CLOCK:
LEVEL? -> :SEARCH:SPI:ANALYZE:SETUP:
CLOCK:LEVEL 1.000E+00
:SEARch:SPI:ANALyze:SETup:CLOCk:
POLarity
Function Sets the edge of the clock channel to be analyzed
of the SPI analysis conditions or queries the
current setting.
Syntax :SEARch:SPI:ANALyze:SETup:CLOCk:
POLarity {RISE|FALL}
:SEARch:SPI:ANALyze:SETup:CLOCk:
POLarity?
Example :SEARCH:SPI:ANALYZE:SETUP:CLOCK:
POLARITY RISE
:SEARCH:SPI:ANALYZE:SETUP:CLOCK:
POLARITY? -> :SEARCH:SPI:ANALYZE:
SETUP:CLOCK:POLARITY RISE
:SEARch:SPI:ANALyze:SETup:CS?
Function Queries all settings related to the CS signal of the
SPI analysis conditions.
Syntax :SEARch:SPI:ANALyze:SETup:CS?
Example :SEARCH:SPI:ANALYZE:SETUP:CS? ->
:SEARCH:SPI:ANALYZE:SETUP:CS:CHANNEL4:
MODE 1;LEVEL 1.000E+00;:SEARCH:SPI:
ANALYZE:SETUP:CS:ESTATE LOW
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel4?
Function Queries all settings related to CH4 of the SPI
analysis CS signal conditions.
Syntax :SEARch:SPI:ANALyze:SETup:CS:
CHANnel4?
Example :SEARCH:SPI:ANALYZE:SETUP:CS:
CHANNEL4? -> :SEARCH:SPI:ANALYZE:
SETUP:CS:CHANNEL4:MODE 1;
LEVEL 1.000E+00
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel4:LEVel
Function Sets the level of CH4 of the SPI analysis CS
signal conditions.
Syntax :SEARch:SPI:ANALyze:SETup:CS:
CHANnel4:LEVel {<voltage>}
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel4:LEVel?
<voltage>=8 divisions within the
screen (0.01 division steps).
Example :SEARCH:SPI:ANALYZE:SETUP:CS:
CHANNEL4:LEVEL 1V
:SEARCH:SPI:ANALYZE:SETUP:CS:
CHANNEL4:LEVEL? -> :SEARCH:SPI:
ANALYZE:SETUP:CS:CHANNEL4:
LEVEL 1.000E+00
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel4:MODE
Function Sets whether to handle CH4 as a CS signal (ON/
OFF) in the SPI analysis conditions or queries the
current setting.
Syntax :SEARch:SPI:ANALyze:SETup:CS:
CHANnel4:MODE {<Boolean>}
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel4:MODE?
Example :SEARCH:SPI:ANALYZE:SETUP:CS:
CHANNEL4:MODE ON
:SEARCH:SPI:ANALYZE:SETUP:CS:
CHANNEL4:MODE? -> :SEARCH:SPI:
ANALYZE:SETUP:CS:CHANNEL4:MODE 1

Related product manuals