ESCAPE TO MEASURE can be used at any stage to abort operation. WARNING: If the HOLD function is activated the instrument returns with the question HOLD (flashing);
answer YES or NO or MODE again to return to measurement.
IM 12D7B2-E-H
16
TEMP = Temperature compensation function NaCl = Automatic temperature compensation The setting of the temperature coefficient is done
by adjusting a conductivity value, at process tem-
perature, to the conductivity value that was meas-
ured at reference temperature
4-4.Temperature compensation
1. Access temperature
compensation routine
2. Select automatic NaCl
temperature compensation
3 Select temperature coefficient
compensation
Access the maintenance menu
Select temperature compensation
function
*
Select temperature compensation
routine
Move pointer to TEMP
NO
YES
Select automatic compensation
Change displayed selection
Confirm selection
NO
YES
NOTE:
1. If NaCl is chosen the temperature compensa-
tion will be according to NaCI solutions.
2. If *NaCl and *T.C. do not appear, this indicates
that Matrix compensation is selected in service
code 13.
Adjustment of the conductivity value
Select digit to adjust
Adjust manual temperature
Confirm adjusted value
ENT
L
L
Error E2 indicates that the temperature coefficient
is less than-10%/°C or more than +10%/°C.
NOTE:
Direct setting of the compensation factor is possible.
Refer to §7-12.
MODE