IM 12D7B2-E-H
Remove cover by releasing 4 screws. Display will show
*
0% Display will show
*
100%
12
4-2. Output range adjustment
1. Access output 2. Adjust low span value 3. Adjust high span value
OUTPUT
SET HOLD
TEMPERATURE
SERVICE
YES
NO
YES
NO
MODE
.
*
OUTPUT
SET HOLD
TEMP.
SERVICE
ENT
µS/cm
ENT
OUTPUT
SET HOLD
TEMP.
SERVICE
ENT
mS/cm
ENT
Access commissioning menu
Adjust value for low span
Select digit to adjust
or decimal point and units
Adjust digit
Confirm adjusted value
ENT
L
*
Return to measurement
select output function
Select OUTPUT
YES
NOTE:
If the output is commissioned to be via a
table this whole item will be skipped! Default
is a linear range.
L
Adjust value for high span
Select digit to adjust
or decimal point units
Adjust digit
Confirm adjusted value
ENT
L
L
MODE
ESCAPE TO MEASURE can be used at any stage to abort operation. WARNING: If the HOLD function is activated the instrument returns with the question HOLD (flashing);
answer YES or NO or MODE again to return to measurement.