IM 12D08B02-01E
Parameter setting 5-17
Code Display Function Function detail X Y Z Default values
User interface
50 *RET. Auto return Auto return to measuring mode Off 0
Auto return to measuring mode On 1 1 On
51 Not used
52 *PASS Passcode Maintenance passcode Off 0 0.0.0 Off
Note # = 0 - 9, where Maintenance passcode On #
Commissioning passcode Off 0 Off
1=111, 2=333, 3=777 Commissioning passcode On #
4=888, 5=123, 6=957 Service passcode Off 0 Off
7=331, 8=546, 9=847 Service passcode On #
53 *Err.01 Error setting Polarization too high Soft/Hard 0/1 1 Hard
*Err.05 Shorted measurement Soft/Hard 0/1 1 Hard
*Err.06 Open measurement Soft/Hard 0/1 1 Hard
*Err.07 Temperature sensor open Soft/Hard 0/1 1 Hard
*Err.08 Temp. sensor shorted Soft/Hard 0/1 1 Hard
*Err.13 USP limit exceeded Soft/Hard 0/1 0 Soft
54 *E5.LIM E5 limit setting Maximum conductivity value 250 mS
(Minimum resistivity value) 0.004 k
*E6.LIM E6 limit setting Minimum conductivity value 1.000 S
(Maximum resistivity value) 1.000 M
55 *% Display mA in w/w% mA-range displayed in w/w% off 0 0 Off
mA-range displayed in w/w% on 1
*0% Set 0% output value in w/w%
*100% Set 100% output value in w/w%
56 *DISP. Display resolution Auto ranging display 0 0 Auto
Display fixed to X.XXX S/cm or M·cm 1
Display fixed to XX.XX S/cm or M·cm 2 (2)
Display fixed to XXX.X S/cm or M·cm 3
Display fixed to X.XXX mS/cm or k·cm 4
Display fixed to XX.XX mS/cm or k·cm 5
Display fixed to XXX.X mS/cm or k·cm 6
Display fixed to XXXX mS/cm or k·cm 7
57 *USP. USP setting Disable the E13 (USP limit exceeded) 0 0 Off
Enable the E13 (USP limit exceeded) 1
58-59 Not used