IM 12D08B02-01E
Calibration 6-3
6-3. Calibration with HOLD active
HOLD
ENT
ENT
>
Select the flashing digit with the > key.
Increase its value by pressing the key
>
After the indication is stable, set the value
using the >,,ENT key.
Put the sensor in standard
solution. Press YES.
When the correct value is displayed,
press ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press
YES.
HOLD will be displayed. Press NO to turn off
HOLD and return to the measuring mode.
NOYES
NOYES
NOYES
HOLD
MODE
NOYES
YES
HOLD
ENT
NO MODEYES
MODE
MEASURE
CAL
DISPLAY
HOLD
Press the MODE key.
The legend CALIB
appears, and the YES/NO
key prompt flags flash.
HOLD
HOLD
ENT
HOLD
HOLD