EasyManuals Logo
Home>YOKOGAWA>Transmitter>SC202G

YOKOGAWA SC202G User Manual

YOKOGAWA SC202G
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
IM 12D08B02-01E
6-2 Calibration
6-2. Calibration procedure
ENT
ENT
>
Select the flashing digit with the > key.
Increase its value by pressing the key
>
After the indication is stable, set the value
using the >,,ENT key.
Put the sensor in standard
solution. Press YES.
When the correct value is displayed,
press ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press YES.
NOYES
NOYES
MODE
NOYES
YES
ENT
NO MODEYES
MODE
MEASURE
DISPLAY
CAL
HOLD
Press the MODE key.
The legend CALIB
appears, and the YES/NO
key prompt flags flash.
ENT
The cell constant is automatically updated after the calibration and the new
value can be read on the display as described in section 4.5.
The calculation is as follows: Cell constant in /cm= (Conductivity of calibration solution in mS/cm) x
(Cell resistance in kOhm)
Comparing this calibrated cell constant with the initial nominal cell constant in service code 03
gives a good indication of the stability of the sensor. If the calibrated cell constant differs more than
20% from the nominal cell constant error E3 is displayed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA SC202G and is the answer not in the manual?

YOKOGAWA SC202G Specifications

General IconGeneral
BrandYOKOGAWA
ModelSC202G
CategoryTransmitter
LanguageEnglish

Related product manuals