8. ERRORS AND COUNTERMEASURES
WITH DISPLAY UNIT
8.1 When Error Is Large or Unstable Is Output
F0801.ai
No
Yes
If a display unit is attached, check
the error code and alarm display.
Connect by HART communication
and check self-diagnostics.
Read Section 6.3 "Self-Diagnostic
(Error Code List)" and check for
recovery measures.
Configure the parameters
correctly.
Is a faulty area found
with self-diagnosis?
Are the parameters
configured correctly at
operating conditions?
Ensure the straight lengths of
pipeline, read Chapter 3
"INSTALLATION".
Replace the gaskets, read
Chapter 3 "INSTALLATION".
This case is due to coating of stuff
vortex shedder bar.
Check the pipinginner surfaces.
This is due to cavitation.
Check for presence of cavitation
P = 2.7 x Δ P + 1.3 x P0 ..... (7)
P: Line pressure (kPa absolute) at
distance of 2D to 7D from
flowmeter on downstream side
ΔP: Pressure loss (kPa)
P0: Saturated steam pressure of
liquid at operating conditions
(kPa absolute)
Is straight lengths
of pipeline stable?
Does the value in
"K34:Vortex frequency"
undergo a lot of
changes?
Is there a snapping
sound from the
flowmeter ?
Does gaskets protrudes?
Is it properly grounded?
Ground properly.
No
Yes
No
Yes
Yes
No
Yes
Yes
No
No
Yes
Normal
No
Check standards.
Yes
<8. Errors and Countermeasures For Display Unit>
43
IM 01F07A01-01EN
Errors and Countermeasures For Display Unit
8

Table of Contents

Other manuals for YOKOGAWA VY Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA VY Series and is the answer not in the manual?

YOKOGAWA VY Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelVY Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals