EasyManua.ls Logo

YOKOGAWA VY Series - Zero Indication Issues

YOKOGAWA VY Series
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.2 Indication Drops to Zero at Certain Times
When this problem occurs, deterioration of sensor sensitivity and turbulence of uid ow due to coating on the
vortex shedder bar and owmeter inner tube are suspected as the cause.
How to cope with this problem
1) Remove the vortex shedder bar from the body according to "Vortex Shedder Removal" in the Maintenance Manual
listed in Table 1.1 Related Documents.
2) If there is coating on the inner tube of the owmeter, remove the owmeter body from adjacent pipes and clean it.
<8. Errors and Countermeasures For Display Unit>
44
IM 01F07A01-01EN

Table of Contents

Other manuals for YOKOGAWA VY Series

Related product manuals