EasyManua.ls Logo

Zeizz Primo Star - Installing;Removing the Mirror

Zeizz Primo Star
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Primo Star / Primo Star iLED Start-Up and Operation Carl Zeiss
M60-2-0011 v 08/2008 33
En
g
lish
3.5.5 Installing/removing the mirror
The mirror serves to illuminate the specimen if no
power outlet is available.
It can be used only in combination with the
Primo Star Fixed-Köhler Microscope. For this
purpose, it is necessary to remove the condenser
and the condenser carrier insert.
Installing the mirror:
Remove the cover cap (
Fig. 30/6) from the
luminous-field diaphragm (
Fig. 30/5).
Unscrew the stop screw (
Fig. 30/8) and lower
the condenser carrier fully down using the
knurled screw for vertical adjustment.
Using an Allen key, loosen the clamping screw
(
Fig. 30/2) of the condenser and the centering
screws (
Fig. 30/3) of the condenser carrier
(
Fig. 30/4) so that the condenser (Fig. 30/7) can
be removed frontward. Remove the condenser
(
Fig. 30/7).
Unscrew the clamping screw (
Fig. 30/2 or Fig.
31/1) from the condenser carrier insert (
Fig.
30/1 or
Fig. 31/3).
Push the condenser carrier insert backward
against the spring and inclining it remove it
upward from the condenser carrier (
Fig. 30/4).
Insert the mirror (
Fig. 31/5) from top through
the opening of the condenser carrier and place
it onto the mount of the luminous-field
diaphragm (
Fig. 31/2). Take care that the mirror
rests horizontally on the mount.
Rotate and incline the mirror until the daylight is
reflected homogeneously into the light path.
Removing the mirror:
Removal of the mirror is done in reversed order.
Fig. 30 Removing/installing the condenser
Fig. 31 Installing/removing the mirror

Table of Contents

Related product manuals