EasyManua.ls Logo

Zeizz Primo Star - Care and Troubleshooting; Instrument care

Zeizz Primo Star
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Primo Star / Primo Star iLED Care and Troubleshooting Carl Zeiss
M60-2-0011 v 08/2008 35
En
g
lish
Carry out this test until the image is format filling without masking or vignetting.
Retighten the set screw (
Fig. 32/5).
With camera/adapter combinations that have not been recommended expressly by Zeiss, it
may be quite impossible to obtain an unvignetted image.
Installing a digital camera or video camera with C-mount thread
Cameras with C-mount thread are to be connected to the phototube of the microscope by means of the
camera adapter P95-C 2/3" 0.65x or P95-C 1/2" 0.5x (
Fig. 32/9).
Put the camera
(Fig. 32/10) together with the corresponding camera adapter (Fig. 32/9) into the
phototube as far as it will go, align it and fasten it by means of the clamping screw
(Fig. 32/8).
4 CARE AND TROUBLESHOOTING
4.1 Instrument care
Care of the microscope is restricted to the following operations:
Cover the instrument with the dust cover after every use.
Do not install the microscope in a humid room; maximum humidity < 75%.
Cover open tubes with the dust caps.
Remove dust and loose dirt from visible optical surfaces with a brush, blower brush, cotton bud, optics
cleaning tissue, or a cotton cloth.
Remove water-soluble dirt (coffee, cola, etc.) by blowing on it and subsequent wiping it off with a
dust-free cotton cloth or a cloth moistened with water to which you may also add a mild detergent.
Wipe off stubborn oily or fatty dirt (immersion oil, fingerprints) with a cotton bud or a dust-free cotton
cloth and the optics cleaning solution L.
The cleaning solution consists of 90 vol% gasoline and 10 vol% isopropanol (IPA). Its individual
constituents are also known as:
Gasoline: Surgical spirit, petroleum ether
Isopropanol: 2-propanol,
dimethyl carbinol,
2-hydroxypropane
Clean optical surfaces by polishing in circles starting in the middle and moving to the edges using slight
pressure only.
Disconnect the plug-in power unit from the supply line before cleaning it. Make sure that no moisture
gets into the plug-in power unit.
For the use in warm and humid climatic zones, all optical components of the microscope are provided
with protection against fungus attack.

Table of Contents

Related product manuals