EasyManua.ls Logo

Zeizz Primo Star - Changing the LED module in the reflected-light fluorescence illuminator

Zeizz Primo Star
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Primo Star / Primo Star iLED Care and Troubleshooting Carl Zeiss
M60-2-0011 v 08/2008 37
En
g
lish
Problem Cause Remedy
Transmitted-light illuminator of
Primo Star iLED cannot be
switched on.
Transmitted light / reflected light
changeover switch is in reflected light
(Fluorescence) position.
Turn transmitted light / reflected light
changeover switch to transmitted light
(Brightfield) position.
Transmitted light / reflected light
changeover switch is in transmitted
light (Brightfield) position.
Turn transmitted light / reflected light
changeover switch to reflected light
(Fluorescence) position.
Reflected-light illuminator of
Primo Star iLED cannot be
switched on.
Reflected-light illuminator not exactly
enough aligned to microscope stand
or displaced, causing interruption of
power supply by interlock function.
Align reflected-light illuminator exactly
to microscope stand and tighten
clamping screw sufficiently.
4.3 Changing the LED module in the reflected-light fluorescence illuminator
ATTENTION
The LED module in the reflected-light fluorescence illuminator has to be changed in any case
by an authorized service technician.
The Primo Star iLED microscope with reflected-light fluorescence illuminator is provided with
an interlock function to switch off the installed LED of the reflected-light illuminator as soon as
the reflected-light illuminator will be turned with regard to the stand or detached or if the lid
of the reflected-light illuminator will be removed.
Disconnect the reflected-light illuminator from
the power supply.
Detach the tube (see also section
3.5.1). To this
end, loosen the knurled screw, turn the tube
counterclockwise by approximately 90° and lift
it out of the dovetail mount.
Loosen the three fastening screws (
Fig. 33/1) of
the lid of the reflected-light illuminator (2.5 mm
Allan key).
Disconnect the plug-in connector of the LED
module connection cable (
Fig. 34/3) on the
circuit board. To this end, press the safety hook
against the connector to release it and pull out
the connector.
Unscrew the two fastening screws of the LED
holder (
Fig. 34/1) completely. Remove holder
(
Fig. 34/2) and screws.
Remove the LED module (
Fig. 34/4).
Insert the new LED module. Because of limited
space, the flat side of the LED module must
point to the circuit board.
Fig. 33 Lid of reflected-light illuminator

Table of Contents

Related product manuals