- the power supply of the building is designed for the comparatively high (possibly
additional) power input of the unit (see technical data) and the line is sufficiently
protected in accordance with regulations!
Do not connect the unit to the power supply until these requirements have been ful-
filled!
• Before putting the unit into operation, check to make sure that the radiator cable and
the radiator are undamaged and have been connected correctly to the unit!
The device must not be placed in front of a radiator or radiant heater.
The device must not be covered by pillows or blankets while in operation.
The device is not made for outdoor operation.
• Do not treat Patients on metal chairs or metal beds.
• Keep credit cards, camera, mobile phone, car key, USB sticks, memory cards and
other interference sensitive storage media away from the unit!
• Clean and disinfect the unit only when the power supply is deactivated (mains switch
off, power plug disconnected)!
• Clean and disinfect the unit only by means of disinfection by wiping! Disinfecting by
spraying can damage the unit due to penetrating moisture!
• Never clean the unit with abrasives, disinfectants or solvents that could scratch the
housing or damage the unit!
• Never perform service work on your own!
• The device must not be used as a step or a seat!
All service work (safety inspections and repairs) must be performed only by service
technicians who have been authorized by the manufacturer!
During service and maintenance work it is not allowed to treat a patient!