36 MicroPro
© gbo Medizintechnik AG Version 1.4
Electrical test (acc. to IEC 62353)
Device leakage current
Protective conductor resistance
Isolation resistance
!!
Note!
The safety inspections have to be documented in a medical device book in
order to document the results of the inspection.
Warning!
If the unit is not safe for operation, then it must be repaired by the authorized
service personnel.
9.3 Cleaning and disinfection
For cleaning and disinfection of the device and accessories, do not use any agents contain-
ing larger amounts of phenol derivatives, chlorine compounds or peracetic acid. Never use
an abrasive. Fingerprints on the touchscreen can be cleaned with a dry cloth or some iso-
propanol.
Switch off the device at mains switch.
Unplug the mains plug from the socket before cleaning or disinfecting the unit.
Clean and disinfect the unit and its accessories on all external surfaces by wiping
with a slightly damp cleaning cloth. Use a commercially available cleaning or disin-
fection agent that is intended for use in medical facilities.
Wait until the unit is completely dry before operating it again.
!!
Note!
The device is suitable for wipe disinfection. We recommend Bacillol-AF for
the disinfection. (Please follow the recommendations of the manufacturer.)