22 MicroPro
© gbo Medizintechnik AG Version 1.4
6 Start of operation
The unit has been completely assembled in the factory and is ready for use except the
installation of the radiator and the radiator cable.
Warning!
For preventing a change of the radiators directionality it should be handled
gently.
Proceed as follows in order to prepare the unit for operation:
Set the mains switch to the off position.
Make sure that the voltage indicated on the type plate matches the mains voltage of
the building.
Plug the radiator cable into the socket on the back of the unit and screw it.
Insert the radiator cable into the recess at the upper end of the support arm. Subse-
quently fasten the radiator cable with the locking screws.
Connect the free end of the radiator cable to the desired radiator. Therefore the lock-
ing ring of the coupler must be pulled back towards the support arm. Subsequently
insert the radiator connector fully into the coupler and release the locking ring. The
locking ring must spring back to its original position. Push and pull the radiator in the
axial direction of the coupler to ensure that the locking ring has also properly en-
gaged. The radiator is now firmly connected.
Plug the cold appliance plug into the corresponding socket on the rear of the unit.
6.1 Check for operational safety
Warning!
The unit and the radiator must be positioned so that there is no danger of
personal injury! Therefore, you must read and observe the safety instruc-
tions and the list of contraindications before putting the unit into operation!
(See 2 Safety instructions and 11 Contraindications)
Check the condition of the housing and the insulation of the radiator, radiator cable
and the power supply cable. Also make sure that the cables have been routed cor-
rectly.
To avoid risk of electric shock, the MicroPro must only be connected to supply mains
with protective earth.
The mains plug is used as separation measure to the mains. The device must not be
positioned in such a way that the mains plug cannot be reached.