MicroPro 11
© gbo Medizintechnik AG Version 1.4
Do not touch simultaneous the patient and touchable contacts of connectors.
Do not treat parts of the patient’s body containing metallic implants, unless special
techniques are used.
At regular intervals during the treatment, check
- that the unit is functioning properly,
- for moisture development
(perspiration) in the area of the radiator and
- whether the patient feels well
!
After the treatment, ask the patient about his tolerance of the treatment and visit the
treatment environment (doctor or physiotherapist).
During service and maintenance work it is not allowed to treat a patient!
2.3 Device protection
Caution!
Improper installation, operation or maintenance of the microwave therapy
unit may result in malfunctions of this device, other devices or other devices
connected to the patient!
Therefore, observe the following instructions in order to prevent mal-functions:
In order to prevent electromagnetic disturbances, place the unit at least 3 meters from
any other devices! Also make sure that there is sufficient distance between the unit
and power supply or data cables in walls, ceilings and floors, since the electromag-
netic radiation from the unit can pass these essentially without hindrance!
In selecting the location for the unit, make sure that the patient has contact during
the treatment to metal elements (especially if they are earthed), such as heating radi-
ators, metal beds or other earthed devices!
Before connecting the unit, make sure that
- the voltage rating on the rating plate corresponds to the mains voltage.
- the frequency rating on the rating plate corresponds to the system frequency,
- an earthed socket outlet with earthing contact is available for connecting the unit,
- the routing of the power cable from the unit to the socket outlet with earthing
contact does not pose a danger for personnel or the patient and that
The affected parts of the body should be unclothed during treatment, since accumulation of moisture on the
skin or in folds can cause local overheating of the skin. This is especially important in case of clothing made
of moisture-resistant fabric such as silk or synthetic fibers!
The output power must always be set according to the subjective sense of the patient! Particular care is to
be taken with patients who have a reduced capacity for heat perception!