Carefully inspect the reagents and samples to make sure there is
no insoluble floatage, such as cellulose, fibrin, etc. to avoid the
reagent & sample probe blocking.
EXI1800 series Chemiluminescence Immunoassay Analyzer
uses cuvettes (cuvettes or plastic cups). Please use cuvettes
specified by Zybio, to avoid unexpected effects.
1) Please use the system as per the instructions in the
operation manual. Incorrect use may result in incorrect test
results and may even lead to system damage or physical
injury;
2) The first test item, the item corresponding to the reagent
should be calibrated, then quality control to verify that the
system is working properly;
3) Daily QC procedures are recommended for routine use of
the system to ensure the reliability of test results;
4) Cover the caps of the sample tray and reagent tray before
analysis;
5) Do not touch the sample tray and the reagent sample tray
when they are rotating to prevent scratches;
6) Do not install any hardware or software on this system other
than those specified by Zybio, otherwise may hinder the
normal use of the system. Do not run other software while
using the system;
7) Do not use this system for other purposes. Incorrect use
may cause the system to be infected with a computer virus.
Computer viruses may transmit through USB, program,
network, etc;
8) There are two fuses with specification of T6.3AL250V near