the power socket of your analyzer. If they need to be
replaced, contact the technical service engineers of Zybio or
its local agents.
1) Please follow the instructions in this manual for system
maintenance. Incorrect maintenance methods may lead to
incorrect analysis results, system damage, or physical
injury;
2) Perform calibration analysis after replacing major
components such as PMT, reagent & sample probe, or
plunger pump;
3) If your analyzer is stopped due to malfunction or other
reasons and needs to be repaired or troubleshooting, please
contact the the technical service engineers of Zybio or it
local agents, and at the same time:
⚫ Please take other measures, such as finishing the test
with other instruments or methods, to avoid delay of the
test results;
⚫ Please take out the reagents from your analyzer and
store the reagents separately according to the
instructions in the reagent kit, or store the reagents in
the refrigerator to prevent reagents from deteriorating.
1) Please use completely separated serum samples and urine
samples without suspended solids or fibrin to avoid blocking
the reagent & sample probe, thus affecting the accuracy of
the analysis results;
2) The presence of drugs, anticoagulants, preservatives, etc. in
the sample may interfere with analysis results;
3) Lipidemia, jaundice and hemolysis in the sample may affect
the analysis results. Blank analysis of samples is
recommended;
4) Please store the samples correctly. Incorrect sample storage
conditions may affect the accuracy of the analysis results by