Safety Information
18
changing the composition or content of the sample;
5) Do not open the samples for a long time to prevent sample
volatilization, or the analysis results may be affected;
6) This system has specific requirements for sample size. When
sampling, determine the appropriate sample size according to
the relevant instructions in this manual.
1.3.15.Reagents, calibrators and controls
Warning
1) Use appropriate reagent, calibrator and control when using
this system for analysis;
2) Please select the appropriate reagents to match this system.
If it is not possible to determine whether the reagent is
appropriate, consult the technical service engineers of Zybio
or its agents;
3) Refer to the reagents instructions for use and storage of the
reagents, calibrators and controls;
4) Inappropriate storage of reagents, calibrator and control may
result in failure to obtain correct test results even within their
shelf life;
5) Perform calibration analysis after changing reagents. Correct
analytical results may not be obtained without calibration and
QC analysis;
6) Cross contamination of reagents during testing may affect test
results. Consult the technical service engineers of Zybio or
agents for reagent cross-contamination information;
7) Verify that the reagents are positioned correctly before
analysis, otherwise the correct results cannot be obtained.
1.3.16.Disposal of your analyzer
Warning
Some substances from scrapped analyzers are regulated by the
contamination regulations. Please adhere to local waste disposal
standards for disposal of scrapped analyzers.