Basic Operating Methods
48
Parameters
Meaning
Operation
No.
Times of calibration test, the newest calibration record is
display at the top
No operation
required
*
The current used calibration fitted concentration
No operation
required
Lot No.
Batch No. of calibrator
No operation
required
Calibration
Time
The time when calibration test finished
No operation
required
Postpone
Calibration valid date will postpone 28 days
Click to select
Refuse
Refuse the fitted curve with successful calibration, only
one curve is not allowed to be refused
Click to select
Result Mark
Mark of calibration result
No operation
required
Status
Status of reagent calibration, including: calibration
success, calibration failure, calibration expired,
calibration extension, calibration required
No operation
required
Red Curve
Master calibration curve
No operation
required
Green Curve
Calibrated curve and fitted curve used in calculating
concentration
No operation
required
Name
Name of calibrator
No operation
required
Number of
repetitions
Number of repeated tests at calibration point
No operation
required
The First Test
Photon quantity value of the first test at the calibration
point
No operation
required
The Second
Test
Photon quantity value of the second test at the
calibration point
No operation
required