Basic Operating Methods
52
Parameters
Meaning
Operation
Cancel request
Cancel QC request test
Select the control item to be canceled
and click [Cancel Request] directly.
Basic operations
◼ Add new QC
QC information is already available when the master calibration curve is scanned. If
additional controls are needed, they can be added as follows:
1) Click [QC] and enter QC interface;
2) Click the additional QC item in [Item List];
3) Click [Add] to pop up the “Add New QC” interface;
4) Enter the QC name and batch number;
5) Set expiration date;
6) Enter the QC target value and SD value;
7) Select the QC rules to be enabled;
8) Click [Save].
◼ Edit the QC
1) Select the items to be edited in the [Item List];
2) Select the controls to be edited in the [QC List];
3) Click [Edit], the edit interface will pop up, the operation method is the same as “Add
New QC”;
4) Click the [Save] button once finished editing.
◼ Delete QC
1) Select items to be deleted in the [Item List];
2) Select the QC item to be deleted in the [QC List];
3) Click the [Delete] button to pop up the prompt box;
4) Click [OK], deletion succeed.
3.11.1.2. QC request
Before QC request test, it is necessary to confirm that there are 1 or more vacancies in the
sample positions.
1) Click [Sample Request], enter sample request interface;
2) Check if the sample positions are sufficient in [Sample List].