Basic Operating Methods
60
Parameters
Meaning
Operation
samples with the
same test information
Save
Save sample
information
Click to save sample information
Cancel
Not save sample
information
No change
Note:
Sample barcode
3) The barcode is composed of numbers with a maximum length of 18 digits,
which can be used as a sample number;
4) The default format is sample number prefixed with time, which starts with 001
for each sample number;
5) Users can enter freely, and the system automatically converts to the default
format; If the input is out of range and the system will give an error prompt to
ask for re-entry, and the duplicate sample number cannot be set after the
sample position is released, and can only be set after the sample is deleted in
the result interface;
6) The barcode of the calibrator starts with STD for easy identification;
7) The barcode of the control starts with QC for easy identification.
◼ Item status
Mark or color
Item type
Description
Item in yellow font
Unavailable
item
The item is allowed to request but not allowed to
test, reason: reagent balance is 0.
Item in red font
Unavailable
item
The item is allowed to request but not allowed to
test, reason: The reagent needs calibration,
calibration failure or calibration expired.
Item in bright blue
Optional item
The item can be request and test
Item box is in gray
Not optional
item
Unloaded item