Basic Operating Methods
61
Item in bright green
Selected item
The item has been selected
Basic operation
◼ Single sample request
1) [Request]- [New Sample], enter new sample input interface;
2) Select the sample position in the position drop-down box;
3) Select sample test mode, sample type;
4) Input the sample barcode (an analyzer with barcode setting function connected to
LIS, click the [Scan Sample Tray], scan the sample barcode to get sample barcode,
EXI1830 does not have this function, only manual input);
5) Input patient information;
6) Click to select the test item, click once to select, and click again to cancel;
7) Enter the test times for the item, the default is 1 time, and the maximum support is
100 times;
8) Click the [Save], the sample information saved;
9) Automatically jump to the next position, and the next sample can be requested.
◼ Batch request
1) [Request] - [New Sample], enter new sample input interface;
2) Select the sample position in the drop-down box;
3) Select sample test mode and sample type;
4) Input the sample barcode (an analyzer with barcode setting function connected to
LIS, click the [Scan Sample Tray], scan the sample barcode to get sample barcode,
EXI1830 does not have this function, only manual input);
5) Input patient information;
6) Click to select the test item, click once to select, and click again to cancel;
7) Enter the test times for the item, the default is 1 time, and the maximum support is
100 times;
8) Click [Batch], batch number selection drop-down box (1-50) will pop up;
9) Drop down to select number of samples and click [OK], samples of the same test
information will be gathered in batches.
Attention
1) When doing batch request, the sample number of the batch
request and the position on the sample tray are incremented in
order according to the ID and position of the starting sample;
2) The starting sample for a batch request must be an unrequested
sample. If the sequentially increasing sample positions contain
samples with status “Requested”, “Testing”, “Not Completed” or
“Completed”, that sample position cannot be skipped and the user
must select the smallest available continuous sample position to