When the project result is less than the lower limit of the measurement range and
the concentration of the minimum concentration calibrator, it will be displayed as "<the
greater value"; when the project result is greater than the upper limit of the measurement
range and the concentration of the maximum concentration calibrator, it will be displayed
Is "> the smaller of the two".
After the optimization of the setting result display, when the sample luminescence
value is greater than the instrument's highest absolute luminescence value, and the test
calculation fails, the result is displayed as ">the upper limit of the adjusted measurement
range".
The results show that the optimization will not affect the storage, transmission and
backup of the results. The calculation items involved in the project will still use the
original results for calculation, and the calculation items directly display the calculation
results.
Only users with system setting authority can perform optimization settings for result
display.
4.8.2.8.Close batch post-processing
After the last test result of the periodic test is completed, the maintenance process
after the periodic test will be carried out.
Basic operation
Equipped with enhanced cleaning fluid:
1) The default is to open batch post-processing, that is, after the cycle test is
completed, the batch post-processing will be executed automatically
Without enhanced cleaning fluid;
2) The default is to open batch post-processing, that is, after the cycle test is
completed, the batch post-processing will not be executed automatically
Equipped with enhanced cleaning fluid;
3) Turn off the batch post-processing, that is, the cycle test will not automatically
execute the batch post-processing
Without enhanced cleaning fluid;
4) Turn off batch post-processing, that is, after the cycle test is completed, the batch
post-processing will not be executed automatically.
Batch post-processing process:
◼ Check whether the buffer bottle, reaction cup, waste cup box (available not less
than 3), and waste liquid bucket are full. If there is an abnormality, report an
error message and stop the subsequent maintenance process;
◼ Check whether there is an enhanced cleaning solution in the reagent position. If