The minimum default aeration time on the 3M™Steri-Vac™ Sterilizer/Aerator GS Series can be changed for the 38°C and 55°C cycles. Press the minimum aeration time and an iWheel will
appear to adjust the minimum aeration default time. Figure 23illustrates changing the minimum aeration defaulttime.
Sterilizer
22.2
˚
C
04/20/2015
12:38:29
98.0
kPa
Select cycle temperature:
65432
1
7
55 °C
Return
Apply changes
Cycle access level
Touch check box to select cycles listedon start screen.
Touch row to change values.
Operator
Cycle
38 ˚C
55 ˚C
2.0 hrs
2.0 hrs
Minimum aeration
Setup
Preferences Cycle Setup Date-Time Network Site
55 ˚C
Minimum aeration
10
11
12 00
hh:mm
:
3013
14
00
:
12
00
:
12
OK
OK
Cancel
Cancel
Figure23.
Changing the Minimum Aeration
DefaultT
ime
The Site Setup>Setup tab (Figure 24) provides the specific configuration and maintenance details for your site; this is established by authorized 3M service personnel duringinstallation.
Sterilizer
22.2
˚
C
04/20/2015
12:39:27
98.0
kPa
Select cycle temperature:
65432
1
7
55 °C
Return
Apply changes
Ventilation hood
High elevation
Abator always on
One door
Abator mode
Lock out immediate
Abator setting
1 to 1
Next maintenance
10-2015
Reset
Reset
Barcode override count
0
System cycle count
38
Last maintenance
04/20/2015
Maintenance interval
6 months
Setup
Preferences Cycle Setup Date-Time Network Site
Ventilation hood
High elevation
One door
Figure24.
Site Setup>Setup Tab -Options