49
12.8.1. Unloading the 3M™Steri-Vac™ Sterilizer/Aerator GS Series – CycleComplete
Sterilizer
55.0
˚
C
98.0
kPa
Stop
Aeration time elapsed (hh:mm):
Cycle complete
0:00
1:00
12:00
Menu
Menu
04/20/2015
21:04:01
Gas expose time elapsed (hh:mm):
Cycle time remaining (hh:mm):
Cycle: Count:3355
˚
C
Open door Show chart
Figure57.
Cycle CompleteScreen
1. Press the Open Door button on the touch screen (Figure57). On double door units, press the Open Unload Doorbutton. The Load Door cannot be opened after the cycle has successfully
completed on double door units.
2. Allow chamber pressure to equilibrate to room pressure; this typically requires 60-90seconds.
3. Open the GS Series sterilizerdoor.
4. Remove the load from the chamber. Remove the empty 3M™ Steri-Gas™ EO GasCartridge.
5. Press the Stop button to generate a printout of the cycle report (Figure58). The display will return to the mainscreen.
6. Follow the Process Monitoring and Quality Control procedures described in Chapter13.
Sterilizer
22.2
˚
C
98.0
kPa
Stop
Aeration time elapsed (hh:mm):
Cycle complete
0:00
1:00
12:00
Menu
Menu
04/20/2015
21:03:12
Gas expose time elapsed (hh:mm):
Cycle time remaining (hh:mm):
Cycle: Count:3355
˚
C
Open door Show chart
Press Stop to Obtain a
Printout CycleReport
Figure58.
Command to Obtain Cycle
ReportP
rintout