Sterilizer /Aerator GSSeries – Operator’sManual
12. OperatingInstructions
12.1. Starting aCycle
The Operator controls the 3M™ Steri-Vac™ Sterilizer/Aerator GS Series via a touch screen. A fingertip, stylus, ball tip pen or a computer mouse can be safely used to navigate the touch
screens. To use a computer mouse, simply plug the mouse into one of the two USB ports (reference Figures 6 and 7) and the mouse arrow will appear on the touch screen. There are seven (7)
programming steps to start a cycle.
DANGER: To reduce the risks associated with exposure to ethyleneoxide,
inspect the display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in thismanual.
CAUTION: To reduce patient risks associated with exposure to potentially non-sterile devices or inadequate sterilizationprocedures,
inspect the display screen and cycle reports for error codes and listen for audible notifications (if enabled). Always take action for error codes as indicated in thismanual.
1. SelectCycle
The 3M™Steri-Vac™ Sterilizer/Aerator GS Series has two validated preprogrammed cycle options: 38°C (100.4°F) and 55°C (131.0°F). Select the appropriatecycle. On a double door
sterilizer the Unload door must be closed to program a new cycle.
Sterilizer
22.2
˚
C
98.0
kPa
38 °C
Open door
Menu
Menu
Select cycle temperature:
55 °C
04/20/2015
05:19:51
65432
1
7
Figure33.
Step One: SelectCycle