57
15. Cautions, Error Messages, andTroubleshooting
DANGER: To reduce the risks associated with exposure to ethyleneoxide:
Never use force to access the inside of the sterilizationchamber.
Inspect display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in thismanual.
Call 3M Health Care service personnel or authorized 3M service personnel immediately if there is a failure of the display or backlight and the sterilizer continues tooperate.
WARNING: To reduce the risks associated with fire andexplosion,
inspect the display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in thismanual.
CAUTION: To reduce patient risks associated with exposure to potentially non-sterile devices or inadequate sterilizationprocedures.
inspect the display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in thismanual.
The 3M™Steri-Vac™ Sterilizer/Aerator GS Series is designed with embedded software that automatically controls, monitors and regulates the mechanical functions of the GS Series sterilizer
and the sterilizationprocess.
15.1. CautionMessages
If the 3M™Steri-Vac™ Sterilizer/Aerator GS Series detects a fault condition that does not adversely affect the performance of the sterilization cycle, a caution message will appear on both the
display screen and cycle report to notify the Operator (Figure 64).
The cycle can complete successfully when a caution message is asserted. Document all caution messages per facilityprocedures.
04/20/2015 08:29:56
Error level: C
Stage : EO injection
Info: SystemModeStateMachine
C604-Power interruption
Hide
Clear
Letter ‘C’ indicates message
is a caution. Document per
facility procedures.
Figure64.
Example of a CautionMessage