EasyManua.ls Logo

3M Steri-Vac GS Series - 12.5. Insert 3 M Steri-Gas EO Gas Cartridge, Load Chamber, and Close Door

3M Steri-Vac GS Series
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
36
3M
Steri-Vac
Sterilizer /Aerator GSSeries – Operator’sManual
Sterilizer
22.2
˚
C
98.0
kPa
Open door
Cycle: 55 ˚C
Aeration (hh:mm): 12:00
Load ID: 120413-1
Menu
Menu
Back
Back
Do not scan
Do not scan
Rescan
Rescan
Next
Next
04/20/2015
05:23:46
65442 31 7
Cannot read barcode
Scan 3M
TM
Steri-Gas
TM
EO Gas Cartridge
barcode:
t
Figure39.
Example Display Message for an Invalid
3M™ S
teri-Gas™ EO GasCartridge
If a valid Steri-Gas EO Gas Cartridge has a damaged barcode, the Operator has the option to bypass the scan. However THIS OPTION CAN ONLY BE PERFORMED FIVE TIMES, after which a
3M Health Care service personnel or authorized 3M service personnel must be contacted to re-set the override (Figure40).
Sterilizer
22.2
˚
C
98.0
kPa
Open door
Cycle: 55 ˚C
Aeration (hh:mm): 12:00
Load ID: 120413-1
Menu
Menu
Back
Back
Do not scan
Do not scan
Rescan
Rescan
Next
Next
Scan 3M
TM
Steri-Gas
TM
barcode:
04/20/2015
05:23:50
65442 31 7
There are 5 barcode
overrides remaining
before a service call is
required. Do you want
to override barcode?
Yes
Yes
No
No
Figure40.
Steri-Gas EO Gas Cartridge Barcode
OverrideMessage
If enabled the sterilizer will require a Supervisor PIN to bypass or override scanning of the 3M™ Steri-Gas™ EO Gas Cartridge. Reference the Site Setup Preferences menu (Figure 21).