36
3M
™
Steri-Vac
™
Sterilizer /Aerator GSSeries – Operator’sManual
Sterilizer
22.2
˚
C
98.0
kPa
Open door
Cycle: 55 ˚C
Aeration (hh:mm): 12:00
Load ID: 120413-1
Menu
Menu
Back
Back
Do not scan
Do not scan
Rescan
Rescan
Next
Next
04/20/2015
05:23:46
65442 31 7
Cannot read barcode
Scan 3M
TM
Steri-Gas
TM
EO Gas Cartridge
barcode:
t
Figure39.
Example Display Message for an Invalid
3M™ S
teri-Gas™ EO GasCartridge
If a valid Steri-Gas EO Gas Cartridge has a damaged barcode, the Operator has the option to bypass the scan. However THIS OPTION CAN ONLY BE PERFORMED FIVE TIMES, after which a
3M Health Care service personnel or authorized 3M service personnel must be contacted to re-set the override (Figure40).
Sterilizer
22.2
˚
C
98.0
kPa
Open door
Cycle: 55 ˚C
Aeration (hh:mm): 12:00
Load ID: 120413-1
Menu
Menu
Back
Back
Do not scan
Do not scan
Rescan
Rescan
Next
Next
Scan 3M
TM
Steri-Gas
TM
barcode:
04/20/2015
05:23:50
65442 31 7
There are 5 barcode
overrides remaining
before a service call is
required. Do you want
to override barcode?
Yes
Yes
No
No
Figure40.
Steri-Gas EO Gas Cartridge Barcode
OverrideMessage
If enabled the sterilizer will require a Supervisor PIN to bypass or override scanning of the 3M™ Steri-Gas™ EO Gas Cartridge. Reference the Site Setup Preferences menu (Figure 21).