Store the comparator limits into a specified ID memory.
Command
M L, 1 2 3 ,+1010,+990
C
L
F
Reply
AK
ID number: Maximum 6 digit
Upper limit: Maximum 9 digit including sign
Lower limit: Maximum 9 digit including sign
Store the total count and number of additions into a specified ID memory.
Command
M A, 1 2 3 ,+6789,
123C
L
F
Reply
AK
ID # U
pp
er Limit Lowe Limit
ID # Total Count # of Additions
ID number: Maximum 6 digits
Total count: Maximum 9 digits including sign
Number of addition: Maximum 8 digits
Request the scale to reply with the F-Function settings.
Command
? F C , 0 5 0 1
C
L
F
Request the setting of
f-05-01
.
Reply
F C , 0 5 0 1 , 0
C
L
F
f-05-01
=
0
Command
? F C
C
L
F
Request to reply with the setting of the next F-Function.
Reply
F C , 0 5 0 2 , 0
C
L
F
f-05-02 = 0
Store the specified F-Function setting value
Command
F C, 0 5 0 3,+1234
C
L
F
Set “1234” as the Upper limit.
F-# Settin
g
value
F-# Settin
g
value
F-Function number: 4 digits
Setting value: Maximum 8 digits including sign.
Reply
AK
Command
O N
C
L
F
Start with new settings.
Reply
AK
Having finished the "FC" command, send the "ON" command to start the scale with
new settings. The scale replies <AK> (06H) and starts.